Specifications
*Note: In compliance with international standards, all Parker portable particle counters can
meet the ISO Medium test dust standards. The icountLCM’s, in addition to the complete
range of Condition Monitoring products, are capable of achieving certification to ISO
4406:1999 and with traceability to ISO 11171 for SRM 2806, via ISO 11943.
icountLCM20
Portable Particle Counter
Specification
Accessory Kit for icountLCM Classic
Re-chargeable
battery pack
Instruction
manual & bar
code software
Power lead,
printer ribbon,
bar code pen &
weather cover
Batteries &
printer reel
Power supply
Limit socket, fuse
& jack plug
Screwdriver
ParSmart &
cable assy
dimensions in inch (mm)
1.10 (28mm)
4.96 (126mm)
9.80 (249mm)
Description
ABS structural foam and injection moulded case
ABS handheld display
Mechanical composition – Brass,
plated steel, stainless steel and aluminium
Fluorocarbon seals
Perfluoroelastomer seals
Nylon hoses (kevlar braided microbore)
Stainless steel armoured hose ends
1.2m fluid connection hose
Rechargeable battery pack
12Vdc power supply
Fast blow fuse
Unique optical scanning system
Bonded glass optical window enclosed in SS plate
Micron channels analysis (Six)
Analysis range ISO 7 to 22 incl. (NAS 0 to 12)
32 character dot matrix LCD. Alpha numeric keypad
Data retrieval
Calibration to ISO standards*
Viscosity range 2 to 100 cSt. 500 cSt.with SPS
Operating temp. 41°F to 176°F (+5°C to +80°C)
Ambient temp. 41°F to 104°F (+5°C to +40°C)
2 minute test completion time
Memory store – 300 test memory
Battery operated 6 x 1.5 D cells
Phosphate Ester group compatibility
Mineral oil & petroleum based fluid compatibility
Up to 420 bar (6000 psi)
Integral 16 column printer
RS232 to USB computer interface
Astra board case weight – (lb.)
Unit weight – (lb.)
ParSmart software and cable link pack
Weather protector cover
CE certified
Auto logging
LaserCM
(LCM20 20 22)
LaserCM
(LCM20 20 62)
11 11
17.6 17.6
4
1.26 (32mm)