Specifications
47
Quick Connection
Connecting the IOS is quick and reliable. The uid connectors are on the front
panel, with two secure push ttings: .236” diameter (6mm) inlet and .157”
diameter (4mm) outlet/drain. Parker can supply dedicated hoses and ttings for
use with most hydraulic and hydrocarbon uids.
Tough Storm Case
The robust waterproof IP54 (when open) case and fully sealed impact resistant
brushed stainless steel front panel provide excellent protection in the most
demanding of applications. The combined unit weighs under 12.1 lb (5.5kg),
making it an ideal ‘ rst use’ diagnostic service tool.
Fast Contamination Detection
The IOS provides fast detection of the presence of contaminants, with the
results being shown on the front panel mounted, high visibility OLED digital
display. This provides easy identi cation of uid condition, showing measured
codes, the sizes per channel in microns, the user de nable limits and moisture
sensor readings as a % of relative humidity.
Complies with the Latest Standards
The IOS is designed in accordance with the latest global standards including:
Long Life Remote Operation
The IOS uses a long life regulated 12 Vdc power supply, with an M12, 4 pin
connector, plus a rechargeable NiMH detector battery unit for use onsite or in
remote locations.
Fluid and Pressure Control
The IOS automatically adjusts ow rates, to an optimum level of 60ml/min.
Total ow range is between 40 and 140ml/min, with maximum online operating
pressure being 36 psi (2.5Bar). An optional inlet reduction valve is also available
for high pressure applications.
• CE marking
• EC Declaration of Conformity
• Machinery Directive
• EMC EN61000-6-3:2001
• EMC EN61000-6-2:2001
• EN 61010-1:2001
Results are viewed in the OLED
digital display window
Pressure Reducing Valve (PRV)
A pressure compensated PRV device (Parker Hanni n part
number ACC6NN027) has been developed to enable testing
where ow pressures in the hose exceeds 35 psi (2.5 bar), up
to a maximum of 5000 psi (350 bar).
Connect INLET .236” Ø (6mm) hose.
Manual Connection: Press the
Pressure Reducing Valve rmly into
the INLET port.
Low Pressure Connection
High Pressure Connection
Parameter Value
Working pressure range 0 to 35 psi (2.5 bar)
Working pressure with PRV 35 psi (2.5) to 5000 psi (350 bar)
Working viscosity 1 to 300 cSt