Specifications
Connection Instructions
Single Point Sampler
Features & Benefits
The Single Point Sampler provides a
means to connect an icountLCM20 to a
single pressure test point and balance the
differential pressure across the system,
to provide a controlled flow of oil into the
icountLCM20 and away into a waste oil
receptacle.
Lightweight, compact and easy to use
design
Fingertip operated control valve even at
high pressures
6,000 PSI (420 bar) rated
Facilitates testing from large diameter
pipework
Capability to test up to 500cSt viscosity
oils (pressure permitting)
Pressure compensated flow control
mechanism
Possible to control the valve with the
same level of accuracy whether the device
is operating at high or low pressure
Capable of allowing a flow rate in excess
of 10ml/min when operating at any
viscosity within the product specification
Suitable for fluid temperatures from +5°C to
+80°C (+41°F to +176°F)
High quality polished finish. (stainless
steel/ aircraft grade aluminium)
Capable of working with an icountLCM20
connected into a system via the standard
one metre extension hose kit
Suitable for use with mineral and
biodegradable oils, petroleum based and
phosphate ester fluids
Phosphate ester version utilizes the
5
/8”
BSF HSP style fitting
Designed so that it meets the lowest
possible level of magnetic permeability
Supplied with accessories kit
It will maintain the set flow rate between
upper and lower limits within a 100 bar
inline pressure change
Clear product identification to ensure that
it is connected correctly (i.e. downstream
of the icountLCM20).
1. Ensure valve is closed (A).
2. Connect P2 on icountLCM20 to P2 on Single Point Sampler (SPS) (C).
3. Connect drain line on SPS (D).
4. Connect P1 of icountLCM20 to the system (F).
5. The SPS is ready to operate.
6. Open valve (A) slowly until the oil flows continuously from the drainline
(D) into a reservoir or recepticale (R).
7. Switch on monitor and begin testing.
icountLCM20 Only
Carry out flow test as shown in the manual. If test is showing below Dt
38.5°F (3.6°C) then carry out test as normal. If, however, test is above Dt
38.5°F (3.6°C) then increase oil flow by turning valve (A) anticlockwise and
then carry out flow test. Do this until Dt is below 38.5°F (3.6°C) and carry
out test as normal once achieved.
WARNING! Ensure that SPS valve is closed and icountLCM20 is
connected to the SPS BEFORE connection to system.
icountLCM20
System
Single
point
sampler
Prs.
(F)
(D)
R
P1
P2
(A)
(C)
(B)(E)
24