Specifications

Technical specifications
icountOS ordering information
Feature Specification
Product start-up time 10 seconds minimum
Measurement period Default 30 seconds run time; 15 seconds data logging time
Reporting interval On board data storage every second. Output via RJ45 connection
Principle of operation Laser diode optical detection of actual particulates
International codes range Up to ISO 22 (+/- 1 ISO code) NAS 0-12
Calibration Calibration by recognised online methods confirmed by the relevant ISO procedures.
MTD - via a certified primary ISO 11171 automatic particle detector using ISO 11943 principles.
Particl
e distribution reporting to ISO 4406:1999.
Recalibration and Servicing Recommended every 12 months
Working pressure 2.5-350 bar (35-5000psi) Pressures above 2.5 bar require the use of a Parker Pressure
Reducing Valve (PRV) - ACC6NN027
Working viscosity 1-300 cSt
Flow range through IOS 40-140ml/minute; controlled at 60ml/min by IOS’s internal pump
Fluid connection interface INLET: 6mm push-fit. DRAIN: 4mm push-fit
Ambient storage temperature for unit -40°C to +80°C; -40°F to +176°F
Operating temperature for unit -30°C to +80°C; -22°F to +176°F
Operating humidity range 5%RH to 100%RH
Fluid operating temperature (Oil) +5°C to +80°C ; +41 °F to +176°F
Moisture sensor Linear scale within the range 5%RH to 100%RH
Computer compatibility IP68-rated RJ45 connection that may be connected to a laptop computer’s RJ45 LAN port
using the 2m cable supplied, WiFi capability 20 metres indoors.
Power requirement Regulated power supply supplied with the unit
Certification IP54 rating (unit open) EMC EN61000-6-3:2001
IP67 rating (unit closed) EMC EN61000-6-2:2001
EC Declaration of Conformity EMC EN61010-1:2001
Machinery Directive CE Certified
Part Number Fluid Type Calibration Connection Option
IOS1210EUR Mineral MTD Offline None
IOS1220EUR Mineral MTD On line None
IOS1211EUR Mineral MTD Offline WiFi
IOS1221EUR Mineral MTD On line WiFi
Description Part number
Hose Kit Bag
(includes one power pack,
RJ45 patch cable and low
pressure hose connectors)
ACC6NN029UK
ACC6NN029EUR
ACC6N N029US
Pressure Reducing Valve (PRV) ACC6NN027
(Standard with IOS 1220)
Po
wer Pack (UK 2m cable) ACC6NN040
Power Pack (EUR 2m cable) ACC6NN041
Power Pack (US 2m cable) ACC6NN042
Description Part number
RJ45 LAN Connector Cable ACC6NN028
Carry Strap ACC6NN030
The Carry Strap option MUST
be selected at the time of
placing the IOS order.
Low Pressure Hoses
(4mm and 6mm)
ACC6NN031
High Pressure Hose Assembly ACC6NN034
(Standard with IOS 1220)
Verification Fluid SER.MISC.067
What is included?
Ofine IOS 1210 EUR/UK/US
1 x IOS Oil Sampler Unit
+ 1 x Power Supply
+ 1 x RJ45 LAN Cable
+ Low Pressure Hoses
Online IOS 1220 EUR/UK/US
1 x IOS Oil Sampler Unit
+ 1 x Power Supply
+ 1 x RJ45 LAN Cable
+ 1 x Low Pressure Hose
+ 1 x Pressure Release Valve
+ 1 x High Pressure Hose
Dimensions Ordering Information
Accessory Part Numbers
Important Information
WARNING – USER –RESPONSIBILITY - FAILURE OR IMPROPER SELECTION OR IMPROPER USE OF THE PRODUCTS DESCRIBED HEREIN OR RELATED ITEMS CAN CAUSE
DEATH, PERSONAL INJURY AND PROPERTY DAMAGE.
- This document and other information from Parker-Hannifin Corporation, its subsidiaries and authorized distributors provide product or system options for further investigation
by users having technical expertise.
- The user, through their own analysis and testing, is solely responsible for making the final selection of the system and components and assuring that all performance,
endurance, maintenance, safety and warning requirements of the applications are met.
- The user must analyse all aspects of the application, follow applicable industry standards, and follow the information concerning the product in the current product
catalogue and in any other materials provided from Parker or its subsidiaries or authorised distributors.
- To the extent that Parker or its subsidiaries or authorized distributors provide component or system options based upon data or specifications provided by the user,
the user is responsible for determining that such data and specifications are suitable and sufficient for all applications and reasonably foreseeable uses of the components
or systems. The operation of the products described here in is subject to the operating and safety procedures details of which are available upon request.
Sales conditions
The items described in this document are available for sale by Parker Hannifin Corporation, its subsidiaries or its authorized distributors. Any sale contract entered into by Parker
will be governed by the provisions stated in Parker’s standard terms and conditions of sale (copy available upon request)
12 13