Specification Sheet
Page 3 of 4
Instrument Power
Run time
Instrument continuous run time shall be 24 hours; when there is an active Bluetooth
connection between the detector and ALTAIR Connect, run time shall be 22 hours.
Power supply
Instrument shall be equipped with a rechargeable battery.
Battery life indication
Monitor shall provide icon depicting estimated remaining battery operation time.
Battery icon must always be visible when instrument is powered on.
Charging cradle
Optional charging cradle shall be offered.
Charger input voltages
Chargers shall be available for 110VAC/220VAC and 12-24VDC.
Charging status
Instrument or charging cradle shall provide visual indication of battery charging status.
Calibration
Calibration tools
Unit shall require no special tools for calibration other than calibration cap,
cylinder, regulator, and tubing to supply gas to instrument.
Pushbutton calibration
Calibration shall be easily performed using instrument’s pushbuttons.
Internal instrument access or tools shall not be necessary for calibration.
Calibration cylinder mix
Calibration gas cylinders shall be offered in standard four-gas configurations:
1. Combustible, O
2
, CO and H
2
S
2. Combustible, O
2
, H
2
S and SO
2
3. Combustible, O
2
, CO and NO
2
Instrument shall
be calibrated using one cylinder.
Calibration time
Span calibration shall not exceed 60 seconds.
Automatic calibration
Instrument shall be compatible with optional automated test and calibration
system that is able to store data. External system shall automatically recognize
and
calibrate instrument and retain all calibration records.
Bump test station
Economical bump test station shall be offered to verify field performance.
Test station shall be capable of checking performance of the instrument and
store records.
Sampling Systems
Sampling mode
In addition to standard diffusion mode, instrument must be available with
external powered pump probe option.
Sampling system filters
Pump must contain user-replaceable filters to prevent liquid and dust ingress.
Sample
line length
Instrument must be capable of sample draw from up to 50 ft (15 m) away.
Fluid ingress protection
Sample probe must prevent water and debris from entering instrument.
Data logging (Instrument Data Storage)
Data logging
Instrument must be available with standard data logging.
Event log
Instrument shall record at least 500 events.
Data log capacity
Data log shall record and store data for an average of 50 hours (at one-minute
intervals) without overwriting existing information in normal use.
Gas record content
Data log entries shall contain as minimum date, time and record of peak and
average readings for each gas sensor (oxygen shall be recorded as maximum
and minimum intervals).
Atmospheric record
Instrument shall have provisions to record atmospheric temperature changes.
Record intervals
Time span between data records shall be user-selectable from 15 seconds
to 15 minutes.
Data retention
Instrument data stored in memory shall not be lost or corrupted in event of
sudden instrument power loss.
Activity record
content page
Instrument data log shall record and be capable of reporting significant
instrument events including:
•
Gas and battery alarms
•
Fresh air setups, sensor re-zeroing and calibrations
•
Battery voltage and elapsed run time
•
Reset of PEAK, Min, STEL, and TWA values
Environmental and Durability
Drop test
Can survive 25+ consecutive 4-ft (1.2 m) drops (MIL-STD-810G drop test).
Can survive incidental 25-ft (7.62-m) drop onto concrete.