Technical data

- 54 -
P1.3 USER CALIB. User adjustment (with an external weight)
Adjustment with external weight at any mass in the balance range, however,
not lower than 30% of max range.
CAUTION
Function unavailable in verified balances.
P1.4 CALIB. TEST Adjustment test
This fuction enables comparing the result of internal automatic adjustment
with the value of internal weight saved in balance’s factory parameters.
The process is conducted automatically and its result is shown on the
display (if the balance is connected via RS 232 port with a computer or
printer, the adjustement result will be printed out). Press ESC to go back to
the previous menu.
P1.5 AUTO. CALIB. T. Temperature of automatic adjustment
(balances with internal adjustment)
Determination of factor which signals start of automatic internal adjustment
NONE - no factor will cause start of adjustement
(adjustment automatically turned off)
TEMP. - adjustment triggered by change of temperature
TIME - adjustment triggered by time interval set in P1.6 AUTO CALIB.
C.
BOTH - adjustment triggered by time and temperature.
CAUTION
Function unavailable in non-verified balances.