Technical data

- 53 -
13.2. Menu adjustment settings
P1 CALIBRATION
P1.1 INT.CALIB. | [internal adjustment]
P1.2 EXT.CALIB. | [external adjustment]
P1.3 USER CALIB. | [user adjustment]
P1.4 CALIB. TEST | [adjustment test]
P1.5 AUTO. CAL. T. | NONE/TIME/TEMP./BOTH [automatic
adjustment]
P1.6 AUTO CAL. C. | [time of automatic adjustment]
P1.1 INT. CALIB. – Internal adjustment
Start of internal adjustment process. Adjustment is conducted automatically,
and no user assistance is required. If weighing pan of the balance is loaded,
balance will display a command to unload the weighing pan. If weighing pan is
loaded with a relatively small mass (up to 4% of balance max capacity) then
adjustment process will be carried out automatically, but measuring results
may differ by the amount that was present on balance weighing pan.
or
P1.2 EXT. CALIB. External adjustment (with an external weight)
Adjustment with external weight, which value is saved in factory menu of
balance, function unavailable in verified balances.
CAUTION
Function unavailable in verified balances.