Technical data
- 53 -
13.2. Menu adjustment settings
P1 CALIBRATION
P1.1 INT.CALIB. | [internal adjustment]
P1.2 EXT.CALIB. | [external adjustment]
P1.3 USER CALIB. | [user adjustment]
P1.4 CALIB. TEST | [adjustment test]
P1.5 AUTO. CAL. T. | NONE/TIME/TEMP./BOTH [automatic
adjustment]
P1.6 AUTO CAL. C. | [time of automatic adjustment]
• P1.1 INT. CALIB. – Internal adjustment
Start of internal adjustment process. Adjustment is conducted automatically,
and no user assistance is required. If weighing pan of the balance is loaded,
balance will display a command to unload the weighing pan. If weighing pan is
loaded with a relatively small mass (up to 4% of balance max capacity) then
adjustment process will be carried out automatically, but measuring results
may differ by the amount that was present on balance weighing pan.
or
• P1.2 EXT. CALIB. – External adjustment (with an external weight)
Adjustment with external weight, which value is saved in factory menu of
balance, function unavailable in verified balances.
CAUTION
Function unavailable in verified balances.