Technical data
- 16 -
Balance BPP
• Remove the weighing pan and remove transport protective elements as
shown in the picture above (if these elements are attached).
• After the unpacking of the balance unscrew the bolt (1) which blocks the
calibration mechanism.
• Assemble the weighing pan.
• Make sure the weighing pan is placed properly on the shock absorbers:
the weighing pan cannot touch the housing and should be stable.
• After removing the protective elements and placing the weighing pan, plug
the indicator into the port which is on the back of the housing. Next, plug in
all the additional devices. Only then can the scales be connected to mains
(the power port is located at the back of the balance’s housing).