Technical data

- 141 -
Tare the balance
Format: T CR LF
Accessible responses:
T_A CR LF
T_D CR LF
- command understood and in progress
-
command carried out
T_A CR LF
T_v CR LF
- command understood and in progress
-
command understood but taring range exceeded
T_A CR LF
T_E CR LF
- command understood and in progress
-
time limit exceeded while waiting for stable measurement result
T_I CR LF
- command understood but not accessible at this moment
Give tare value
Format: OT CR LF
Accessible response: OT_TARA CR LF command carried out
Response format:
1 2 3 4-12 13 14 15 16 17 18 19
O T space tare space unit space CR LF
Tare - 9 characters with right justification
Unit - 3 characters with left justification
CAUTION:
Tare value is always given in adjustment unit.
Set tare
Format: UT_TARA CR LF, where TAREtare value
Accessible responses:
UT_OK CR LF
- command carried out
UT_I CR LF
- command understood but not accessible at this moment
ES CR LF
- command not recognised (tare format incorrect)
CAUTION:
Use dot in tare format as decimal point.