Technical data
- 132 -
19.2. Printer
In the submenu <PRINTER> the balance’s user has a possibility of choosing a port
and device, to which the data is sent by pressing button on balance. The
content of the data being transferred is set in parameter <PRINTOUTS/PRINTOUT
GLP>.
Procedure:
− press
button
− enter the menu <P4 DEVICES>
− and then enter in menu group<P4.2 PRINTER>
− set the balance parameters related to the cooperation with the
printer
<P4.2.1 PORT> to which the printer is to be connected
Accessible options:
COM 1 or COM 2 – RS 232 port, to which the printer is connected
USB type A – USB port, to which PCL printer is connected
WIFI –WIFI port, which can send data to special software
manufactured by MRC e.g. WIN measurement, opened on the
computer connected to the balance via WIFI.
USB flash drive –USB port type A, to which external memory flash
drive is connected
USB PC –USB port type B, to which the computer with the special
software by MRC is connected e.g. WIN measurement
Example of the measurement printout is described in the PRINT-OUTS section.
19.3. Bar code reader
The submenu <BAR CODE READER> includes the balance settings for
cooperation with the bar code reader.
Procedure:
− press
button
− enter the menu <P4 DEVICES>
− and then enter in menu group <P4.3 BAR CODE READER>
− set the balance parameters related to the cooperation with the bar
code reader
<P4.3.1 PORT> - choice of the port, to which the bar code reader is
to be connected
o accessible options: NONE, COM 1, COM 2