20. TECHNICAL PARAMETERS 
20.1. MGB series scales 
  Calibration    internal 
  display    5,7” touch panel 
  Interface    2 x USB, 2 x RS 232, Ethernet 
  Operating temperature  + 18
o
 C - + 30
o
 C 
  supply    feeder 100V-240V , 50-60 Hz / 15V DC 
  sensitivity drifter   2 ppm/°C in +15 ° ÷ +35°C t emperature 
  Max humidity    80 % 
  Min humidity    35 % 
MGB-2  MGB-5  MGB-11  MGB-21  MGB-31 
Max capacity  2g  5g  11g  21g  31g 
Elementary unit  0,1µg  1µg 
Tare range  -2g  - 5g  - 11g  - 21g  - 31g 
Repeatability  0,25µg  2µg  3µg  5µg  5µg 
Linearity 
±0,3µg  ±3µg  ±5µg  ±10µg  ±10µg 
Pan 
∅ 20  ∅ 30 
MGB-5/F  MGB-21/P 
Max capacity  5g  21g 
Elementary unit  1µg  1µg 
Tare range  - 5g  - 21g 
Repeatability  2µg  2µg 
Linearity 
±3µg  ±10µg 
Pan 
∅ 100  ∅ 20 