System information
- 75 -
Only those buttons /symbols/ functions are described below, that did not appear In
„Weighing” mode.
After activating PERCENTAGE DEVIATIONS
application, default symbols enabling immediate
operating start are displayed:
− tare setting
− Reference mass
− Reference masses base
− set 100%
− calibration
Reference mass entering
After pressing the symbol, mass should be entered, e.g. 350.0 g, that will be
accepted as 100% pattern. On its base following samples inspection process
will be conducted. For the mass decimal places number should be given.
PROCEDURE:
o Press Quick access button
o Reference mass will be displayed
o Press the mass content field and
enter new value In dialogue
window
o Accept it with OK button.
o Choose decimal places number.
o Pressing the button results In
automatic accepting the value.
Checkweighing
Two checkweighing thresholds should be given as [%] values.
Weighed masses tolerance range is set,
that is permissible area within which fed
mass is accepted as a correct one.
Tolerance range can be asymmetric,
e.g.
− Top threshold: +5%
− Bottom threshold: - 2%
The values can be visible In information
field if feeding is chosen for them
Dosing
Target mass should be given; it is to be reached while comparing following
loadings to reference mass.
Example:
Reference mass =100g
Target mass = 60%
----------------------------------------------
Real weighed mass = 62,0g
Displayed value 62,000%
Bar chart is green – sample is in permissible field, because target
mass is achieved.