System information

- 61 -
18.8.3. ACAI function – Automatic Accuracy Correction
ACAI function is responsible for jest automatic correction of detail’s individual
mass during counting. It results in higher counting precision standard –
individual mass counted from larger sample size is a representative value.
First ACAI function should be
activated.
PROCEDURE:
o At any time of operation Press
grey information field.
o Screen with internal menu will be
displayed
o Rewind menu to Patterns ACAI
field
o press ACAI button
o choose YES option from dialogue window
ACAI function operating rules:
1. items number (after adding)
located on pan must be higher
than it was bafore
2. items number (after adding)
located on pan must be lower
that double amount that was
visible on display before adding
3. current items number must be included In allowance field ± 0,3 from total
value
4. result must be stable.
ACAI function active state is
indicated by black mark next to
items number. If In information
field „Pattern mass” information
visibility is declared, after
conditions 1-4 meeting automatic
individual mass updating will take
place.
The new value can be saved In „Patterns base” using quick access button.