System information

- 39 -
Refreshing
The function defines the number of shown momentary
measurements while weighing series. All measurements can be
shown, then many transitory states can be seen. Another extreme
situation is showing as few as possible measurements number. For
example only 3 measurements can be seen before stability reaching.
Autozero Autozero
Each measurement series starts with so called,,ZERO”, which is
shown with symbol on scale’s display. In fact ,,ZERO” is an
area usually of 0,25e quantity. Some processes like filling, feeding,
evaporation require only autozero system operation, which can forge
the measuring result. That is caused by very laminar process’
characteristics.
AUTOZERO is active in most of scales. To turn it off, choose NO
option. After moving on to weighing, ,,ZERO” symbol will not be seen
on main display.
Last figure
The function has 2 setting options: sound: YES or sound: NO.
Suitable value should be chosen. After active area choosing, display
returns to previous window loading changes automatically.
18.4.4. Buttons operating
The menu contains settings connected with SCALE REACTION on user’s action. The
settings contain:
Left sensor
Choose one of operations to be performed by scale’s left zooming sensor.
Accessible options:
o off
o print
o tare
o door opening /option accessible for XAB-Y, MGB scales
o door closing / option accessible for XAB-Y, MGB scales
o door changing / option accessible for XAB-Y, MGB scales
units