System information

- 38 -
18.4.3. Readout
The menu contains information connected with WEIGHING PROCESS such as:
Filter
Before displaying, each measuring signal is electronically
converted In order to reach suitable parametres typical of
stable, that is ready to be read signal.
At certain state, user can influence the conversion range,
choosing suitable FILTER. Range of choice, filter:
o Very fast
o Fasts
o Medium
o Slow
o Very slow
Choosing filter level, real scale’s operating conditions should
be taken into account. For very good conditions, medium or
Fast filter can be set, while for Bad conditions slow or very
slow filter will be suitable.
Caution:
For precise scales very fast –medium range filters are reccomended
For analytic scales and microbalances medium- very slow range filters are
reccomended
Result accepting
decides about measurement result stability symbol displaying
moment; one of 3 result accepting options can be set:
o fast
o fast + precisely / reccomended value/
precisely
Caution:
Stable result reaching Speed depends on filter type and result accepting type.