System information

- 30 -
10.1. Calibration report setting
Scale calibration report procedure enables proving that device is
correctly operating and that measurement series shows real
measured product’s mass.
According to approved system of procedure documentation, user
can use In calibration report following information and symbols:
operating mode type
date
time
user
scale type
scale’s OD
level state
signature
blank line
dashes
nominal mass
current mass
difference
defined code
to define codes to be used In calibration report, use User’s Profile
/Operating Modes / Weighing / ID Definition /
10.2. History of calibration
Programme of scale records all conducted calibration procedures
and saves the information In its protected memory in case of mains
failure. From the menu level it can be specified what kind of
calibration should be placed on the list.
Following types of information can be displayed:
All calibrations
Temperature calibrations
Time calibrations
External calibrations
Internal calibrations
User’s calibrations
Hint: If memory is overloaded (after 50 adjustment procedures), the earliest list’s
record will be removed automatically.
If internal procedures require complete calibrations documentation, lists of records
should be printed and stored.