System information
- 29 -
1 Internal calibration
Pressing the field causes starting internal calibration process. It is
completely automatic with no scale’s operator’s interference; if
pan is loaded, appropriate command to remove loading will be
displayed.
2
External calibration
Pressing the field causes starting calibration with external weight,
whose value is given in manufacturer’s menu. The function is not
available for verified scales. Weight value change is not possible.
3 User calibration
Pressing the field causes starting calibration with optional weight ,
which should be given before calibration starting. The function is
not available for verified scales.
4 Calibration test
Comparing internal calibration weight to its value recorded in
scale’s memory. It results in deviation defining scale indication
drift.
5 Weight correcting
possibility of correcting internal calibration weight value; function
not available for verified scales
6 Automatic calibration
definition of factor deciding about internal automatic calibration
starting; options:
no – no calibration
time – calibration by time set in <time cal.Aut.> parameter
temperature – calibration by environment temperature changes
both – calibration considering time and temperature changes
7 Automatic calibration time
defining time intervals between following automatic calibration
procedures – setting range: 1-12 hours
8 Report print
defining if a report should be generated and printed after finishing
adjusting or test procedure
9 Report setting
Calibration content report setting
10 Calibration history
History of calibration procedures