7.4. Automatic tare using for weighing
Operation rules:
each first measurement after stable state achieving is tarred. Net message will be
displayed. Net mass can be defined now.
PROCEDURE
o At any time of operating press grey
information field
o Internal menu connected with the
function will be displayed
o Rewind menu to AUTOMATIC
TARE field and press its area
Caution:
In „Tare“ menu, minimal mass value above
which automatic tare starts operating,
should be defined.
o Set automatic tare ACTIVITY as
YES
o minimal mass, above which tare
starts operating, should be defined
Caution:
Automatic tare does not have its own Quick Access Button. If Active=YES is set up, tare
starts operating. Active=NO setting turns tare off.
7.5. Digital tare
Digital tare is entered with QUICK ACCESS Button. After pressing the button, tare
value should be entered with accuracy not higher than that of the scale. Entered
tare preceded with MINUS sign is automatically displayed.