Instruction manual
70 
1 - 3  4  5  6 7 8 - 16 17 18 - 20 21 22 
Command 
space 
stability 
marker 
space 
mark 
mass 
space 
unit 
CR 
LF 
Command  – 1 ÷ 3 marks 
stability marker – [space] if stable 
[?] if unstable 
[^] if there is an error of exceeding the range on + 
[v] if there is an error of exceeding the range on - 
mark  – [space] for positive values or [-] for negative values 
mass  – 9 marks alignment to the right 
unit  – 3 marks alignment to the left 
16.  WEIGHING LOADS UNDER BALANCE 
In standard analytical and precision balances can weigh load on suspension. 
To use this function: 
•  Remove plastic plug in basic of balance 
•  There is suspension in the basic of balance. It’s installed for good. 
•  Install hook to hang load (hook isn’t standard equipment for balance), weigh 
load on hook. 
Attention: 
•  Suspension cannot be turned around, move or manipulate. Mechanism of 
balance can be damaged this way. 
•  Mass all additional elements as balance, string should be settled to zero 
by pressing key Esc/TARE. 