Instruction manual
31 
10.4. Printout of all balance parameters through RS 232 port 
After pressing F ke
y, balance parameters will be send through RS 232 port. 
Return to weighing 
(see - 7.1.2. – return to weighing). 
11.  BALANCE CALIBRATION 
As the accele
ration value due to gravity is not the same at every location on earth, 
each balance must be coordinated – in compliance with the underlying physical 
weighing principle – to the existing acceleration due to gravity at its place of location 
(only if the balance has not already been adjusted to the location in the factory). 
This adjustment process must be carried out for the location in the factory). This 
adjustment process must be carried out for the commissioning, after each change 
of location as well as in case of fluctuating environment temperature. It is also 
recommend to adjust the balance periodically during weighing operation in order to 
obtain exact measured values. 
To ensure high precision of weighing corrective factor in relation to standard mass 
must be noted in the balance memory periodically – it is the balance calibration. 
Calibration should be performed: 
- Before the weighing, 
-  When long breaks are between following measure series 
-  When temperature inside the balance changes more than: 3° C. 
Kind of calibration: 
- Internal automatic calibration 
* started if temperature changes 
* started if the time changes 