Setup guide
GRAVIMETRIC GATEWAY AND ETHERNET COMMUNICATIONS
29
4. Open a Terminal program such as HyperTerminal. (Click the Start button,
choose Run and type HYPERTRM.EXE, click ok)
5. Start a new connection and choose to connect using the COM port you have
connected the GW21W-MAXI to on your PC.
6. Set the baud rate to 9600 bps for console port communication. Do not set it to
115200 bps, which may be specified in the GW21W-MAXI’s instructions. This
is incorrect.
7. Set the Data bits to 8, Parity to None, Stop bits to 1, and Flow control to
None.
8. Use the following instructions to configure the GW21W-MAXI.
9. After finishing the console settings, power off the GW21W-MAXI, put SW1
and SW2 back to the previous settings.
Entering Console Mode
1. Verify your GW21W-MAXI. You should have SW1 in the OFF position and SW2 in
the ON position. The mini din to RS-232 DB-9 cable should be connected to the
GW21W-MAXI’s din port, The DB-9 side of the cable connected to a null modem,
then to your computer’s COM port. If this is correct proceed with step 2.
2. With HyperTerminal open and connect to your COM port at 9600 bps, power on the
GW21W-MAXI.
3. You will be presented with the following output to the HyperTerminal screen:
ABLELink Ethernet-Serial Server
User name:admin
Password:
4. By default there is no password, press enter. You will see:
Login ok
0.Exit 1.Overview 2.Networking 3.Security 4.Com1 5.WLAN
Input choice and enter(0~5):