ioLogik R2140 User’s Manual Modbus/TCP Address Mappings
Address Data Type Description
0x000D 1 word CH1 AI Range
00: +/-150 mV
01: +/-500 mV
02: +/-5V
03: +/-10V
04: 0-20 mA
05: 4-20 mA
Others: return Illegal Data Value
0x000E 1 word CH2 AI Range
00: +/-150 mV
01: +/-500 mV
02: +/-5V
03: +/-10V
04: 0-20 mA
05: 4-20 mA
Others: return Illegal Data Value
0x000F 1 word CH3 AI Range
00: +/-150 mV
01: +/-500 mV
02: +/-5V
03: +/-10V
04: 0-20 mA
05: 4-20 mA
Others: return Illegal Data Value
0x0010 1 word CH4 AI Range
00: +/-150 mV
01: +/-500 mV
02: +/-5V
03: +/-10V
04: 0-20 mA
05: 4-20 mA
Others: return Illegal Data Value
0x0011 1 word CH5 AI Range
00: +/-150 mV
01: +/-500 mV
02: +/-5V
03: +/-10V
04: 0-20 mA
05: 4-20 mA
Others: return Illegal Data Value
0x0012 1 word CH6 AI Range
00: +/-150 mV
01: +/-500 mV
02: +/-5V
03: +/-10V
04: 0-20 mA
05: 4-20 mA
Others: return Illegal Data Value
B-4