User`s guide
“No Configuration File Loaded”
This is usually caused by trying to load a configuration file for one type of
module/system into a different type of module/system.
Verify that the appropriate module has been selected from the Load
{module} from File {filename} in the HP 16500A/B disk operation menu.
Selecting Load {All} will cause incorrect operation when loading most
preprocessor interface configuration files.
See Also Chapter 1 describes how to load configuration files.
“Selected File is Incompatible”
This occurs when you try to load a configuration file for the wrong module.
Ensure that you are loading the appropriate configuration file for your logic
analyzer.
“Slow or Missing Clock”
This error message might occur if the logic analyzer cards are not firmly
seated in the HP 16500A/B or HP 16501A frame. Ensure that the cards
are firmly seated.
This error might occur if the target system is not running properly.
Ensure that the target system is on and operating properly.
If the error message persists, check that the logic analyzer pods are
connected to the proper connectors on the preprocessor interface. See
chapter 1 to determine the proper connections.
For HP 16510A Logic Analyzers, check the preprocessor interface power
fuse in the logic analyzer.
If You Have a Problem
“No Configuration File Loaded”
MC68306 Preprocessor A–13