User guide

Veloce 50 Operating Manual
14
7.2.2 HOW TO CLEAN THE TRANSDUCERS AND CONDITIONER CONNECTORS
Transducer connector
STEP 1 STEP 2
FISO Technologies recommends cleaning the
connector via the use of FISO Fiber Optic
Connector Cleaner. To do so, ensure a new
cleaning surface by advancing cloth tape. Tear of
f
excess tape as required.
Softly press connector and expose it to the cloth
with motion illustrated in above figure. Up to 6
connectors may be cleaned before advancing
tape.
Conditioner internal connector
STEP 1
STEP 2
Provided in the cleaning kit, swab sticks are
required to insure a proper cleaning technique.
Please do not use cotton swabs. Swabs sticks are
available at most electronic store. Take a swab
and pour a few drops of isopropanol
on it (avoid
acetone).
While rotating, smoothly
insert the swab in the
internal connector of your conditioner. Then rotate
the swab in order to swamp the entire internal
connector surface. Repeat step #1-2, 3 times and
always use new swabs sticks. To preven
t
scratches on the connector, always use soft tip
swab stick