Instruction Manual
The Moticon SCIENCE Sensor Insole surface can be cleaned with desinfectants and a samp soft wipe.
Reprocessing
The sensor insole is not a sterile product.
Do not spray the desinfectant on skin or feet. Pay attention to the safety instructions provided
by the desinfectant manufacturer.
Only use desinfectant based on the active ingredients didecyldimethylammonium chloride
and ethanol.
If worn by different wearers, the sensor insole must be reprocessed before it can be reused.
Cleaning and disinfection procedure:
1. Remove the battery lid
2. Remove coin cell batteries and attach battery lid
3. Spray disinfectant on the bottom side and
on top side of sensor insole
3
4. Clean the bottom side and the top side of
sensor insole with damp soft wipe, then dry
with a clean soft wipe
4
1
2
16
Moticon SCIENCE Insole Instruction Manual 03.00.00