Operation Manual

37
4 After 3 seconds, the microscope automatically memorizes the optimum brightness.
5 Set the optimum brightness of the reflected light path for the mounted objectives by repeating
the above for each of them.
3.16.3 Setting the fluorescence turret channel manager
Ɣ It is effective when the PA53 FS6 revolving fluorescence turret with sensor and the LED lamp
housing Lumos are used.
Ɣ The desired wavelength and intensity can be set per each combination of filter cube (for lower
excited light noise or even different wavelength light intensity for multi-band filter cube) and
objective lens.
Once the wavelength of Lumos for filter cubes is set, selecting an channel of the turret and
objective lens adjusts the optimum Lumos color and intensity automatically.
1 Switch to reflected fluorescent lighting path, (so that the light intensity indicator on the front of
frame is lit in blue).
2 Engage a filter cube in the turret and bring the specimen in focus.
3 Press brightness plus and brightness decrease simultaneously to turn off/on the desired
wavelength (Lumos pad U channel for UV excite lightB channel for blue excite light
G channel for green excite light).
4 After 3 seconds, the microscope automatically memorizes the optimum wavelength of Lumos
for the filter cube.
5 Set the optimum wavelength of Lumos for the mounted filter cubes in turret by repeating the
above for each of them.
Control Pad
(Fig. 37)