3.9 Installing the slider with the filter holder
Ɣ Insert the slider into thefilterslider slot of the
fluorescence attachment. (Fig. 27)
(Fig. 27)
3.10 Shutter Slider (Fig. 28-1)
Ɣ Blocking the optical path of the Epi-fluorescence by sliding the shutter to the left. (Fig. 28-2)
(Fig. 28-1) (Fig. 28-2)
3.11 Installing the Filter Cubes
COMMENT: WE HAVE 5 FILTER CUBES AND ONE EMPTY POSITION FOR BF TL
CAUTION: TURN OFF THE LAMP BEFORE SWITCHING OR INSERTING NEW FILTER CUBES!
Caution
Ɣ Always keep 5five filter cubes in the turret. Do not leave any of the filter slots empty as strong
lightmay enter and damage the user’seyes when shifting betweenexcitation methods.
Ɣ If less than five filter cubes are required for microscopy. Use a “DIA-ILL” dummy cube to ensure
that the open space in the turretis filled.