Operation Manual
9
Ɣ For changing the bulb of the light source mercury lamp the usage of suitable protection equipment
(Solid gloves and Face protection mask) is mandatory
Ɣ WARNING: UV emitted from this product. Avoid eye and skin exposure to unshielded product.
Ɣ WARNING: Possibly hazardous optical radiation emitted from this product. Do not look at
operating lamp. Eye injury may result.
Ɣ CAUTION: IR emitted from this product. Avoid eye exposure. Use appropriate shielding or eye
protection.
Device Safety and EMC
PA53 FS6 is designed, manufactured and inspected in accordance with the IEC 61010-1:2010/EN
61010-1:2010, GB4793.1-2007, GB4793.9-2013and YY 0648-2008safety standards for electrical
measurement, control, adjustment and laboratory device.
Electromagnetic is compatible to the standard of IEC 61326-1:2012/ EN 61326-1:2013/ IEC
61326-2-6:2012/ EN 61326-2-6:2013 and EN 55011 level
RS232port shall only be connected to certified computer according to IEC/ANSI/UL 62368-1
Transporting, unpacking, storage of the Instrument
Please observe and follow the safety notes for transportation, unpacking and storage
given in this document:
Ɣ The microscope is delivered as set, packed into commercial standard plastic and cardboard
packaging; Please re-use the original packaging only for any transportation.
Ɣ It is advised to keep and use the original packaging for longer storage or return to the manufacturer
to avoid losing the warranty.
Ɣ At receiving and unpacking the equipment, please verify that all parts specified on the delivery note
are present.
Ɣ Keep Transport and storage temperatures as specified in this Manual.
Ɣ Set the microscope up on a stable worktable with solid and smooth top surface suitable for
Instrument use.
Ɣ Do not touch optical surfaces.