User Manual
v
Caution(s)
• To prevent possible damage to the keypad, do not use sharp or hard objects to depress keys, only use
fingertips.
• Do not attempt to clean the device or patient cables by submersing into a liquid, autoclaving, or steam
cleaning.
• Wipe the exterior surface of the device and patient cables with a sterilizing disinfectant, and then dry with
a clean cloth.
• Conductive parts of the patient cable, electrodes and associated Type CF connections, including the neutral
conductor of the patient cable and electrode, should not come into contact with other conductive parts,
including earth ground.
• The Ambulatory X12+ and Patient Cable should be cleaned between each use.
• Do not pull or stretch patient cables as this could result in mechanical and/or electrical failures. Patient
cables should be stored after forming them into a loose loop.
• The Ambulatory X12+ will only work with electrocardiographs that are equipped with the appropriate
option.
• No user serviceable parts are inside. Any
modification of this device may alter defibrillator protection.
Any modification to any part of this device is to be performed by qualified service personnel only.
• The following equipment may cause interference with the RF channel: microwave ovens, diathermy units
with LANs (spread spectrum), amateur radios and government radar.
• To prevent possible damage to the device during transport and storage (while in original packaging) the
following environmental conditions must be adhered to:
Ambient Temperature Range:
-20°C to 65°C (-4°F to 149°F)
Relative Humidity Range:
5% to 95% (non-condensing)
Atmosphere Pressure:
700 hPa to 1060 hPa
• Allow the device to stabilize within its intended operating environment for a minimum of two hours prior
to use. The allowable operating environment is as follows:
Ambient Temperature Range:
0°C to 45°C (32°F to 113°F)
Relative Humidity Range:
5% to 95% (non-condensing)
Atmosphere Pressure:
700 hPa to 1060 hPa