Table X-2 Guidance and Manufacturer’s Declaration: Electromagnetic Immunity
The X12+ Transmitter is intended for use in the electromagnetic environment specified below. The customer or the
user of the X12+ Transmitter should assure that it is used in such an environment.
Emissions Test
Compliance
Compliance
Level
Electromagnetic Environment:
guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
+/- 6 kV contact
+/- 8 kV air
+/- 6 kV contact
+/- 8 kV air
Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical Fast
transient/burst
IEC 61000-4-4
+/- 2 kV for
power supply line
+/- 1 kV for
input/output lines
Not Applicable
Surge
IEC 61000-4-5
+/- 1 kV
differential mode
+/- 2 kV common
mode
Not Applicable
Voltage dips,
short
interruptions and
voltage variations
on
power supply
input
lines
IEC 61000-4-11
<5% UT
(>95% dip in UT)
for 0.5 cycle
40% UT
(60% dip in UT)
for 5 cycles
Not Applicable
Power frequency
(50./60 Hz)
magnetic field
3 A/m
3 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
Note: UT is the a.c. mains voltage prior to application of the test level.