User Manual
SECTION 2
11
LED Indicators
WAM Status
LED
+ Audio
MODE
Unpaired with Device or
OUT OF RANGE
GREEN OFF
YELLOW OFF
Audio tick at 2Hz,
Starting 2 sec after unit
is turned ON.
If the unit is unpaired
(meaning does not have
a host assigned) the
status is permanent
until pairing occurs.
If paired but OUT OF
RANGE the GREEN and
YELLOW LED are OFF,
the audio is ON but if
the unit goes back in
range, audio will stop
and GREEN or YELLOW
led will resume status as
normal.
Lead Fault
YELLOW solid (unless
Low battery )
GREEN OFF
One or more leads is
not connected properly.
ALL OK
GREEN ON solid
YELLOW OFF
No lead fail condition is
detected, Battery is OK.
Collecting 10 seconds of
data…..
GREEN ON solid
YELLOW OFF
Audio at 2 Hz until 10
seconds of data are
acquired and the
electrodes can be
disconnected.
The WAM is collecting a
10 second ECG.
Low Battery
Blinking LED (Yellow or
Green depending on
lead fault status)
The WAM has detected
a low battery condition
please replace the
battery within 15
minutes.
Very Low Battery
GREEN OFF
YELLOW OFF
¼ sec audio on, then
units turns off
The WAM has detected
the very low battery
status and will turn OFF.