User Manual
SECTION 2
9
Patient ECG Hookup
Skin Preparation
Skin preparation is important to perform before electrode attachment to help ensure good signal quality when
transmitting patient data. Poor skin-electrode contact may cause noise or artifact which can affect the analysis of the
ECG data. Low amplitude signals may also be the result of poor skin-electrode contact.
To prepare the skin
1. Identify the electrode sites on the torso by referring to Positioning the Electrodes.
2. Remove any hair from the electrode sites using a razor.
3. Wipe oils from the electrode sites with an alcohol prep pad.
4. Remove any dead skin from the electrode sites with an abrasive pad. Two to three moderate rubs at each site
should be sufficient.
Positioning the Electrodes for 10-wire (12-lead) Hookup
Limb
Electrodes
Placement
AAMI
IEC
RA
R
Right arm lead should be
placed on the right deltoid,
forearm, or wrist.
LA
L
Left arm lead should be
placed on the left deltoid,
forearm, or wrist.
RL
N
Right leg lead should be
placed on the right thigh or
ankle.
LL
F
Left leg should be placed
on the left thigh or ankle.
Precordial
Electrode
Placement
AAMI
IEC
V1
C1
Fourth intercostal space at the right sternal border
V2
C2
Fourth intercostal space at the left sternal border
V3
C3
Midway between V2 and V4
V4
C4
Fifth intercostal space at the left midclavicular line
V5
C5
Anterior axillary line on the same horizontal level as V4
V6
C6
Mid-axillary line on the same horizontal level as V4 and V5