User Manual
SECTION 2
8
Turning the WAM On
The WAM will power up as soon as a battery has been inserted into the battery compartment and the cover applied.
If the WAM was turned off after its last use, the user can power the WAM on by pressing the On /Off button on the
front side of the cable yoke:
Turning the WAM Off
The user has two options to power the WAM off:
1. Remove the battery cover, or
2. Press the On/Off button
a. The LED display will turn off
b. An audible tone will sound indicating power off and RF disconnect
Connecting the WAM to the ELI 350 (Pairing Mode)
To pair the unit with the ELI 350:
1. Enter WAM pairing mode on the ELI 350.
2. Bring the WAM very close to the ELI 350.
3. Ensure the unit is off
4. Enter pairing mode on the WAM by holding the ECG 12 Lead and Rhythm buttons
simultaneously and pressing the On/Off button.
5. Wait a few seconds,
6. If the WAM beeps three times it is permanently paired with the ELI 350 in pairing mode. Pairing
will be retained until this process is repeated.
7. If three beeps are not heard and the green and yellow LEDs are toggling intermittently, pairing
has failed. Please try steps 1-6 again.