Mooshimeter Cheat Sheet Version 0 Jan.
Measurement Overview High Voltage Input: Modes: Ranges: ● DC ● ±1.2V ● AC ● ±60V ● ±600V 10MΩ input impedance ± 1.2V range is floating Current Input: Modes: Ranges: ● DC ● ±1A ● AC ● ±2.5A ● ±10A Connected to C (Common) input by current sense resistor and fuse. Aux. Input: Modes: V Ranges: ● DC Voltage ● ±100mV ● AC Voltage ● ±250mV ● Resistance ● ±1.2V ● Diode Drop Ω Ranges: ● 1kΩ ● 2.5kΩ ● 10kΩ ● 1MΩ Common Input: ● 2.
Protection Overview High Voltage Input: 10MΩ input impedance Tested to CATIII standard – 4kV surge Current Input: Factory installed fuse rated to 10kA breaking at 600V Polycarbonate Case: Tested to 4kV sustained from any terminal. Aux. Input: PTC and clamp protected to 600V overload.
Scan Page Scan results: These are meters that were detected in the scan. Tap one to connect to it. If the connection is successful, you will be taken to the Meter View Page. Scan button: Scan for Mooshimeters within range. Results will appear in the list as they are detected. Scan ends after 10 seconds.
Meter View Page Preferences Opens the Preferences Dialog Channel 1 Control See “Channel Control” Channel 2 Control See “Channel Control” Sampling Control See “Sampling Control” Rotate to Landscape Will enter Graph View
Channel Control Measurement Type On the V and A inputs, tap to cycle between DC and AC. Input Select Tap to change which input port is being used. This label corresponds to the marking on the case. On the Ω input cycles between: ● DC Volts ● AC Volts ● Resistance ● Diode Drop Measurement Range When manual control enabled, tap to cycle between available ranges Units Auto-Range Enable Displays the units of the Tap to toggle reading.
Sampling Control Sampling Rate The frequency at which the ADC pulls samples. The green button enables manual control. Sample Buffer Depth The number of samples the meter takes to perform DC and AC calculations The green button enables manual control. Note that AC calculations are inaccurate at buffer depths below 128. Logging Enable Tap to enable logging. With logging enabled, if an SD card is installed, samples will be saved to it when the phone is disconnected.
Graph View – Trend Mode All settings from the Meter View are carried over to Graph View Auto-ranging is disabled in Graph View. To change settings, switch back to Meter View by turning the phone to portait orientation. Channel 1: Red Pause/Play Button Starts and stops data flow to the chart. In iOS version this is done by tapping the background.
Graph Control Trend to Buffer Mode Toggle Trend Mode displays data aggregated slowly over time Buffer view grabs a single sample buffer from the meter for closer analysis (useful for AC) Channel Toggles Toggles the display of Channel 1 and 2.
Graph View – Buffer Mode Refresh Samples and download a new buffer from the meter