Specifications
9
AUTOMATIC OPERATION:
Once the desired test parameters are set up, the user simply selects the “start” button to
begin the tests. All the parameters are shown in the automatic screen. As soon as the
start button is pushed the screen switches to the auto test screen and proceeds to run
the prescribed number of decays and balance (only if balance is set to run via the setup).
The unit moves automatically from one test to another until it has completed the
programmed sequence. At the end of the tests the screen changes to display, individual
decay times, average decay time, balance results, including +/− peaks and average
voltage, as well as date, time, temperature, humidity, group/location. From this point you
can select to run another test sequence or return to the automatic screen.
As in the manual mode, the group/location button will allow you to select the appropriate
label for the ionizer under test.
Programming features:
Start voltage – Same as manual
Charge Differential – Same as manual
Stop voltage – Same as manual
Test Start Delay – Same as manual
Max Decay Time – Same as manual
Balance Duration – Same as manual
Decay Cycle – Range from 1 to 10, adjustable in increments of 1. This is the number of
+ and – decays the unit will run in an automatic sequence.
Decay Sequence – Select either the decay sequence of + − + − or + + − − for the number
of cycles selected in Decay Cycle.
Cycle Delay – Range from 2sec to 15sec, adjustable in 1sec increments. This is the
amount of time from the finish of the last decay cycle to the start of the next.
Balance (Y/N) – Select whether or not you want a balance test to automatically run at the
end of the decay cycle.
Continuous – This feature allows you to perform a continuous series of tests on a
selectable time basis. For example, you want to run a series of decay and balance every
hour for the next day. From the automatic screen select CONT, then select the desired
test time interval from 1 minute to 24 hours. Once you have programmed the time, press
EXIT and you will advance to the next screen. From this screen you have the option of
pressing CANCEL or START. The START key begins the default automatic test
sequence and will repeat that test sequence at the time interval selected. This
continuous testing will continue until you stop the tests or the memory becomes full.