Appendix A: Performance Specification
The specification tables use the following abbreviations:
Performance Requirement (Req). All performance requirements in
the specification tables are identified with
Req,
and can be assured
by completing the Performance Check Procedure located in the
service manual. Allow a warm-up time of 20 minutes.
Performance requirements in the electrical specifications are valid
over an ambient temperature range of +20° C to +30° C. The
Performance Requirement tolerances listed in the Electrical
Specification are doubled over the temperature range of 0° C to
+40° C, unless otherwise specified. Test equipment used to verify
Performance Requirements must be calibrated and working within
the limits specified in the service manual Equipment Required List.
Reference Information
.
Information that amplifies a performance
requirement or is of special importance is indicated by
REF. There is
no need to check these items to a specific tolerance.
Safety Standards
ANSI/ISA S82Safety Standard for Electrical and Electronic Test,
Measuring, Controlling, and Related Equipment.
CAN/CSA C22.2 No. 1010.1-92Safety Requirements for
Electrical Equipment for Measurement, Control, and Laboratory
Use.
IEC1010-1Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use (1990).
UL3111Standard for Electrical Measuring and Test Equipment.