Calibration of pendulum weight
Although all Ergometers are calibrated at the factory
the user may wish to verify this by performing a
mechanical scale calibration. If so, please do the
following.
Remove the cover from the flywheel. Loosen the
balancing spring from the brake belt. Check that the
0-index of the scale is in line with the index of the
pendulum weight. Adjust if needed. Note: Wait until
the flywheel is not moving any longer.
Connect a known weight, e.g. 4 kg (Art. No: 9000-
221) where the balancing spring was placed. Note:
The weight should not be lighter than 3 kg, due to the
possibility of inferior accuracy.
When correctly set, it should be possible to read this
weight from the corresponding place on the meter
panel.
Should there be a deviation, adjust the pendulum
to the correct position on the scale by means of the
adjusting weight. In order to change the position of
the adjusting weight, loosen the lock screw(1) of the
weight.
Should the index of pendulum weight be too low,
move the adjusting weight(2) upwards into the weight.
Should the index be too high the adjusting weight is
moved downwards and locked in the new position.
This procedure is repeated until the correct reading
is achieved.
Check the calibration of the pendulum weight once a
year or when needed.
Assemble the front cover again.
Zero adjustment of meter panel
Move the pendulum to 4 kp and keep it there for
a few seconds. Check that the belt is loosened. If
adjustment is necessary, loosen first the lock nut(3)
and then change the position of the meter panel(5),
so that the board will have its 0-index in line with
the index of the weight. Tighten the lock nut after the
adjustment.
Fig: Calibration
1) Lock screw
2) Adjusting weight
3) Lock nut
4) Brake belt
5) Meter panel
6) Weight