Troubleshooting guide
Monark 839 E
9
Connection to PC
To connect a PC to the ergometer, use a 0-modem
cable (RS232) with a 9-pin D-sub female at each end.
If no RS232 serial port is available on the PC use a
USB serial adaptor to connect to an USB port.
Follow these steps to install the USB adaptor drivers:
Locate the USB adaptor. 1.
Inside the USB adaptor packaging, there is a mini-2.
CD.
Insert the mini-CD into the CD-ROM drive and 3.
install the driver software. If there is no CD drive
on your computer, driver software is available for
down from the website (http://www.vscom.de/
USB-CD).
Finalize the driver software installation by 4.
inserting the USB adaptor.
Install the Monark Software disk. 5.
Connect the USB Adapter to the Serial Cable and 6.
proceed with testing.
To control the ergometer use the PC software supplied
with the ergometer and which is common to the
models 831 E och 939 E, or other PC software made
for the Ergomedic 839 E.
From software version MEC3V11R14 and later
settings can be made from a PC in terminal mode
if the terminal is not available. Set PC in terminal
mode. A terminal emulator is normally available in
i.e. Windows under Accessories/Communication.
In terminal mode do the following settings:
- 9600 baud
- 8 data bit
- 1 stop bit
- no parity
- no flow control
- set terminal emulation to VT100
- set the COM port number.
A USB serial converter is automatically
assigned to a COM port number by Windows.
This number is indicated under Startmenu /
Settings / Control Panel / System Hardware.
Connect Ergometer and PC with the 0-modem cable
(normally used for the ergometer terminal).
Turn on power to the ergometer. The ergometer is
now checking what type of device is connected. When
finished a message appears on the PC screen.
Common commands:
Calibration: Type: cali• [enter] and follow the
instructions on the screen.
Setting to control ergometer from other external •
ECG devices: Siemens Megacart ECG device:
Type: env cmdtype=1[enter]
Setting to control ergometer from external ECG •
devices: Various ECG devices: (most common
setting) type: env cmdtype=2[enter]
To go back to control the ergometer with the terminal
type: env cmdtype=[enter]
A lot of other settings can be made. For more
information about this please see Technical Reference
Manual MEC3V11Rn.