Setup guide

31
Range of Motion
1. Select Sit & Reach
Click Calibrate.
The portable Sit & Reach station
calibrates the exact same as the 700
system.
2. Pull the handle slowly out to the foot
plate and click the Next> button.
3. Pull the handle out exactly 12” (30cm
if using metric units) PAST the
footplate*, then click Next>.
The correct calibration constant is
between 0.35 and 0.75.
* You must have a 12” ruler when
performing this calibration.
Note: It is easier and typically more accurate
to use two people when calibrating the Range
of Motion.
WARNING: Use care when pulling out and returning the handle. Do not pull the handle out too far and
when a client has finished the test be careful they do not let go of the handle and let it snap back. This is a
delicate measurement device.
12”