Molift Mover 300 L EN - User manual BM06201 Rev.
Molift Smart 300 L / www.etac.com Table of contents Table of contents ............................................................................................................................... 2 English manual .................................................................................................................................. 4 Important ............................................................................................................................................
Molift Smart 300 L / www.etac.com Charging ...........................................................................................................................................12 Transfer ............................................................................................................................................13 Using slings........................................................................................................................................13 Lifting and lowering .
Molift Smart 300 L / www.etac.com English manual Important This user manual contains important safety instructions and information regarding the use of the lifter and accessories. In this manual the user is the person being lifted. The carer is the person operating the lifter. Warning! This symbol indicates important information related to safety.
Molift Smart 300 L / www.etac.
Molift Smart 300 L / www.etac.com General CE declaration The product, and its related accessories, described in this instruction for use (IFU), is in compliance with the regulation (EU) 2017/745 of 5. April 2017 – as a medical device, risk class I. The product is manufactured and tested according to standards IEC 60601-1, IEC 606011-2 and EN/ISO 10535:2006.
Molift Smart 300 L / www.etac.com Technical data Safe Working Load (SWL): Maximum A-weighted sound power level: 300 kg (661 lbs) LWA = 63,2 dB Weight of unit: Operating forces button: Excl. battery and sling bar: 36 kg/79,3 lbs Chassis: 19 kg/41,8 lbs Lifting column: 17 kg/37,4 lbs Battery: 1,6 kg/3,7 lbs 4-point medium sling bar: 2,5 kg/7,3 lbs Buttons on handset: 3.
Molift Smart 300 L / www.etac.com Assembly Battery and hand control Charge the battery. The battery must not be in the battery holder during assembly! 1 2 3 Lifting column 1. Connect hand control to bottom of battery holder. 2. Insert battery. 3. Connect hand control wire to push handle so it does not fall to the floor when dropped. Checklist after assembly Place lifting column in chassis. Make sure that the red arrow on the lower part of the column corresponds with the edge of chassis.
Molift Smart 300 L / www.etac.com How to use Molift Mover 300 L General safety precautions Safe Working Load Only use accessories and slings that are adjusted to fit the user, type of disability, size, weight and type of transfer. The lifter is marked with Safe Working Load (SWL) The lowest SWL shall always be used if safe working load (SWL) differs between lifter, sling bar and body support unit. Working pause ratio/Duty Cycle.
Molift Smart 300 L / www.etac.com Safety devices Emergency stop / emergency lowering Molift Mover 300 L is equipped with several safety devices, which are intended to prevent damage or injuries to personnel and equipment in the case of incorrect use. The lifter has an overload sensor preventing the lifter to be operated if the load exceeds SWL. Manual emergency lowering The lifting arm is hinged to prevent the arm and sling bar from squeezing the user.
Molift Smart 300 L / www.etac.com Electronics Transport and operating conditions Battery and service indicator Transport and storage The lifter has two indicator lights on the battery holder. For long time storage it is recommended that the emergency stop button is activated (pushed in). The lifter can be stored and transported under temperatures between -25 - 70 °C. Service indicator Battery indicator Operating The lifter is designed for use at standard room temperatures (+5 to +40°C).
Molift Smart 300 L / www.etac.com Charging Battery indicator (LED) Battery charger Plug (disconnecting device) Battery Charger The battery charger can be used as a tabletop charger or mounted to a vertical wall surface close to a power outlet, minimum 120 cm above the floor. The charger has to be used outside the patient environment. The charger must be placed or installed in a way that makes it easy to disconnect the plug. Take care not to damage the cable.
Molift Smart 300 L / www.etac.com Transfer Plan the lifting operation in advance to ensure that it is as safe and smooth as possible. Remember to work ergonomically. Assess the risks and take notes. The carer is responsible for the safety of the user. Molift Mover 300 L can be used with 4-point sling bar for different types of transfers; sitting, recumbent or ambulating. The sling bar’s hook design prevents sling to be inadvertently detached when mounted correct.
Molift Smart 300 L / www.etac.com Supporting arms (optional) Height of supporting arms: 90 - 150 cm (adjustable) The lifter can be used for gait training with supporting arms. Supporting arms - operating Place the bracket ② for supporting arms in the desired height. Turn lock handle ① clockwise to tighten bracket in position. 1 Supporting arms - general information 2 The supporting arms only function as support to the user.
Molift Smart 300 L / www.etac.com Position lock handle in suitable position to avoid obstructing the operation of the lifter. When needed to release locking arms, turn lock handle ① counterclockwise. 1 To retract the supporting arms, swing the supporting arms upwards 270 degr. Using as walking aid Read user manual for the ambulating vest on how to apply and use it. Ensure supporting arms are in the correct height for supporting the user.
Molift Smart 300 L / www.etac.com Transfer Lowering into sitting Try to position the user as low as possible (preferably resting the feet on the chassis) when transferring the lifter with a suspended user. In this way the centre of gravity will be low and reduce the risk of instability.
Molift Smart 300 L / www.etac.com Accessories Recommended optional equipment and accessories for Molift Mover 300 L. Sling bars and sling bars with scale Molift Mover 300 L can be used with a 4 point sling bar. Sling bars are available with or without Molift Scale. Molift Scale is approved Class III. The scale can also calculate BMI (Body Mass Index). 4-point sling bar Slings Etac supplies a wide selection of slings for different types of transfers.
Molift Smart 300 L / www.etac.com Maintenance Cleaning and disinfection Reconditioning Clean on a regular basis. Clean surfaces with a damp cloth using an appropriate pH-neutral detergent. Do not use solvents or strong liquids, this may damage surfaces on the lifter. For disinfection when needed; use isopropyl alcohol. Avoid abrasive cleaning products. Remove hair and pile from the castors and verify that the castors rotates freely.
Molift Smart 300 L / www.etac.com Troubleshooting Symptom Possible cause / action The lifting column is wobbly The lifting column is not correctly positioned in the chassis or the locking handle is not fully locked. Assemble column and chassis properly. The lifter moves unevenly on a plane surface The castors move unevenly because of pile and dust. Clean the castors. Remove hair and dirt. The lifting arm or adjustable legs moves Short circuit between hand control and lifter. uncontrolled.
Etac A/S Parallelvej 3 DK-8751 Gedved www.etac.