Use Instructions
If even after the instructed actions a problem/warning message reappears,
please disable the device from use and contact the manufacturer for service.
The functionality of the device may have been compromised due to an
internal or external factor (e.g., electromagnetic disturbances).
Description Required action
Desired lens is not available in
treatment planning
Please go to my settings tab and tap My lenses row. All available lenses are listed
and you can select which lenses you want to have visible in treatment planning.
Unselect obsolete lenses and select desired lenses.
Spot size altered
Warning is displayed if spot size is changed after a treatment is initialized. Warning
popup displays the required spot size and current spot size. Adjust the SLA spot
until the current spot size matches the required value and the warning is cleared.
The treatment can then be continued.
Unfortunately
this version of the
application is too old.
iPad software needs to be updated. Device cannot be used before the update is
completed.
You do not have
access rights to device
EIDXXXXXX.
The ML6710i is not defined under your user account. Please connect to a device
you have access to.
Device calibration
expires in XX days
Calibration is about to expire. Please schedule maintenance with the manufacturer.
Device calibration
expired.
Calibration has expired. Please schedule maintenance with the manufacturer.
Device is automatically disabled after 30 days calibration overdue.
Connection to laser is
lost.
iPad is not able to connect to laser. Please check your iPad settings to ensure
Bluetooth connection is in use.
Battery level low. iPad battery is running low. Please recharge it.
Laser stop pressed
LASER STOP button pressed. Please deactivate LASER STOP by pressing the laser
stop button again.
Interlock is
disconnected.
Interlock is not detected. Please check that the interlock is properly plugged in. If
the problem persists, please contact technical support.
Optical fiber is
disconnected .
SLA fiber is not detected. You must properly connect the SLA fiber to the device. If
the fiber is connected and the error message still appears:
Check that the fiber port on the back panel of the device appear clean i.e. no big
particles inside the adapter, and also that the threading is clean so that the nut and
the connector can be twisted to the end.
If the problem persists, please contact technical support.
MANDATORY
Revision 9
59Troubleshooting