Use Instructions

Description Required action
iPad doesn’t connect to the main
unit
Make sure main unit power is on (see section 3.4.5 and that Bluetooth is enabled in
the iPad settings.
Yellow indicator light is blinking
on the device.
Device internal error that prevents normal operation. You must restart the device
by switching the device off and back on. If the problem persists, please contact
technical support.
The device is not visible in the
user interface and all main unit
indicator LEDs are off .
Turn the key switch on the back panel from “O” option to “I. Make sure that power
adapter is properly connected.
Error. Reboot laser warning
message is displayed on the
laser tab.
Device internal error that prevents normal operation. Error should disappear
automatically after a short period and operation can then be continued. If error does
not disappear you need to be reboot the ML6710i main unit (turn power off with the
key, wait 10 seconds and turn power back on).
No internet connection warning
is displayed in login screen
Internet connection is not available. Please check your iPad settings to ensure
network connections are in use and make sure that there’s internet connection
available.
Internet connection is
required for running a
new treatment.
Internet connection is not available. Please check your iPad settings to ensure
network connections are in use and make sure that there’s internet connection
available.
Device temperature
stabilizing.
Device temperature is stabilizing. This is typical if the room temperature is higher
or lower than normally. The message will disappear automatically when target
temperature is reached; this may take a few minutes. If you push OK, the status bar
continues to display the message until target temperature is reached.
Device overheating.
Device temperature is too high. Laser can’t be activated. You must wait until the
device cools down. Make sure the ventilation holes are not blocked and that the
ambient temperature does not exceed specified values.
Noticeable heating of the exterior
of the device
Shutdown the device and wait until device cools down. Make sure the ventilation
holes are not blocked and that the ambient temperature does not exceed specified
values. If the problem persists, please contact technical support.
Dose or irradiance error is visible
in Treatment planning and new
treatment cannot be started.
Please set the dose and irradiance to values requested by the application. The
editing is started by tapping the value.
7. Troubleshooting
7.1 Solving problem and error situations
Problems, warnings and notifications given by the GUI and ML6710i indicators are explained in
Table 13. In case any of these events, follow the instructions given on the GUI and this table.
Table 13. ML6710i warnings and notifications.
ML6710i Instructions For Use
58 Troubleshooting