For loose dirt, try an air duster first. For stains,
wear rubber gloves and wipe the mirror with
a dust-free cloth (e.g., lens tissue) lightly
dampened with 70% ethanol or equivalent
lens cleaning agent. Use gentle gestures from
top to bottom; it is important to move the
cloth only in one direction not to scratch the
delicate optical surface with loose particles.
Cleaning the optical fiber connector
Unless the optical fiber is disconnected,
regular cleaning of the fiber connector is
not required. If you consider cleaning the
fiber connector necessary, use a dust-free
cloth lightly dampened with 70% ethanol or
equivalent lens cleaning agent for cleaning the
connector with one-direction, gentle wipes.
It is recommended to inspect the fiber after
cleaning with a suitable fiber microscope.
The fiber shall be reconnected immediately
after cleaning and inspection to avoid
contamination of the fiber connector. Pay
attention to using the dust caps on the fiber
connectors when disconnecting optical fibers.
Follow the SLA inspection and cleaning protocol to avoid
incorrect beam characteristics and wrong dose.
MANDATORY
Do not touch the optical surfaces of the SLA with bare hands. If the SLA
optics seem damaged or cannot be cleaned with the methods specified
in these Instructions For Use, please contact the manufacturer.
C A U T IO N
Cleaning contact lenses
When using non-disposable contact lenses that are
used from patient to patient, please pay attention to
lens hygiene and maintenance in order to reduce
the risk of cross-contamination via the contact
lens surface in contact with the patients eye.