Use Instructions

Figure 48. Warning displayed when
device calibration due date has passed
and calibration has expired.
Figure 49. Warning displayed when device
calibration is overdue 30 days or more
and device is disabled from use.
After the due date has passed, a warning
message of expired calibration (Figure 48) is
given. Please note that the laser calibration
cannot be guaranteed after calibration overdue.
Therefore, for patient safety, the device is
disabled from use and treatments cannot be
executed after 30 days past the due date.
The remaining number of days before the
disabling is displayed on the pop-up window.
Should the device calibration overdue
pass 30 days or more, the warning
message in Figure 49 is displayed.
No chemicals other than water or 70% ethanol shall be used for cleaning
the device to avoid damage of the device surface and labeling. Do
not pour or spray cleaning agents directly on the device.
C A U T IO N
Cleaning the main unit
The outer casing of the main unit can be
wiped down periodically with a damp cloth.
The device endures normal cleaning for
the entire service life of the device.
Cleaning the SLA
It is recommended to inspect the SLA mirror
regularly and clean when necessary as dust and
other dirt may build up on the mirror surface.
4.4.1 Cleaning
ML6710i Instructions For Use
46 Operating the laser device