Use Instructions

Please note that when the GUI is disconnected some device features that are performed
by the GUI will have limited availability. For example, error messages or elapsed time
are not displayed and ending treatment early on the GUI is not possible. In the case of a
disconnected GUI, treatment may be continued with the foot switch pedal until planned
treatment time has elapsed or can be paused by releasing the foot switch pedal.
C A U T IO N
Please check the Troubleshooting section
of this document for more information
on how to proceed in various cases, also
when experiencing unexpected behavior
of the device (e.g., in the software).
4.4 Maintenance
There are no user serviceable parts inside
ML6710i other than changing of the exterior
power cord. Only power cords provided by
the manufacturer (AP0307, length 1.83 m) can
be used to prevent possible problems with
electromagnetic compatibility (EMC). In addition,
you may temporarily replace the protective dust
cap mounted to the rear panel fiber output port
should it become unattached. A spare piece of
an unattached version is provided with the unit.
Figure 47. Reminder of an approaching calibration
due date.
MANDATORY
In case of an unexpected software behaviour or error, see troubleshooting section in
these Instructions For Use.
MANDATORY
Annual maintenance interval shall be followed. Only personnel authorized by the
manufacturer are allowed to carry out maintenance or service on the device.
due date is 60 days or less away. The message
appears each time the GUI is connected
to the device, and the displayed number of
days remaining is updated automatically.
The GUI prompts when the due date for
periodic maintenance is approaching. The
reminder message in Figure 47 is given when
Revision 9
45Operating the laser device