Use Instructions
4.3.3 Interrupted treatment
The ML6710i is able to continue an interrupted
treatment e.g., in case of power failure in the
building. When the device is again powered
and the Bluetooth connection between the
main unit and iPad is re-established, the GUI
shows the message given in Figure 44.
After tapping continue treatment, the
confirmation pop-up in Figure 45 is shown.
All settings for the interrupted treatment
are restored from device memory and
dose value is based on the light dose
remaining at the time of the interruption.
4.3.4 Device warnings and errors
The device reports warnings whenever required.
If a warning is displayed you should try to fix the
problem. In case an error is detected during an active
treatment, the treatment is automatically paused
and a warning message is given on the GUI.
Figure 46 shows an example warning in case
the connection to the GUI/iPad is lost after
starting a treatment. Should this occur, you may
continue the treatment normally by pushing
the foot switch emission pedal as long as the
green READY emission LED is on and status
LED is white. An audio signal is given by the
ML6710i main unit to indicate laser emission.
The blue EMISSION LED is also lit normally. The
GUI automatically attempts to reconnect with
the ML6710i.
Figure 45. Warning popup is displayed in case the
connection between the GUI and laser device is
lost during treatment. Treatment may be continued
with the foot switch.
Figure 46. Confirmation pop-up for continuing an
interrupted treatment.
If after the instructed actions a problem/warning message reappears, please disable the device
from use and contact the manufacturer for service. The functionality of the device may have
been compromised due to an internal or external factor (e.g., electromagnetic disturbances).
MANDATORY
ML6710i Instructions For Use
44 Operating the laser device