Use Instructions
Figure 33. Treatment planning screen. Interactive
fields are identified by the blue text color.
Set the treatment parameters
Adjust the spot size
Start the treatment
Control the laser by using the foot switch.
4.3.1 Treatment planning
Treatment planning screen in Figure 33 allows
you to modify dose and irradiance values
and set the correct contact lens specific
to each patient. The available value range
for each parameter is shown in Table 9.
Treatment duration is calculated automatically
according to dose and irradiance values set.
The treatment parameters are saved per
user. If you logout and re-login the previously
set treatment parameters are restored.
Parameter Available values
Dose 5 – 95 J/cm2
Irradiance 50 – 800 mW/cm2
Contact lens Selection from list
Aiming beam mode On, Off or Pulse
Aiming beam brightness 1 – 10
Selected spot size 1.0 – 5.0 mm multiplied with the contact lens magnification
Table 8. Available treatment settings.
ML6710i Instructions For Use
38 Operating the laser device