Use Instructions
The middle LED is for indicating device
status as explained in Table 6.
Please note that the green READY indicator
remains on also in the case of an error during
treatment. This is to say that also in the special
case of a disconnected iPad (see section
4.3.4), the treatment is still configured to
the device, but the error situation indicated
by the Yellow LED needs to be cleared
before continuing treatment is possible.
4. Operating the laser device
This section contains instructions on how
to operate the ML6710i. The examples
cover the normal operation of the device
as well as various error situations.
Figure 14. Laser status indicators.
LED state Description
White ON Device is powered on and functioning normally.
White BLINK
iPad (GUI) is disconnected and only limited features are available. If a treatment has been started, it
can be continued with the foot switch (treatment automatically ends after the planned duration).
Yellow ON
Warning – please see instructions given by the GUI and troubleshooting section in these Instructions
for Use in order to solve the issue.
Yellow BLINK Unrecoverable internal error. Device power needs to be cycled off and on to continue.
Red ON
LASER STOP button has been pushed. Treatment cannot be continued before LASER STOP has been
reset by pushing LASER STOP button again or rebooting device (power off and back on).
Red BLINK Unrecoverable internal error. Device power needs to be cycled off and on to continue.
Table 6. Device status indicator.
4.1 User interface on the main unit
4.1.1 ML6710i chassis
On the main unit chassis upper edge there
are LED’s indicating laser and device status.
READY shows green when the device is
ready for use. This means that the treatment
parameters have been confirmed and the laser
emission can be turned on (see subsection 4.3.2)
provided that status LED is in normal state.
EMISSION shows blue whenever the laser
emission is on (treatment running).
ML6710i Instructions For Use
28 Operating the laser device