Use Instructions
Connecting SLA fiber
Before running a treatment, please inspect functionality of the interlock by removing
the interlock connector or by releasing the remote interlock. The change of the interlock
state must be visible on the GUI (see Figure 21). After confirming the functionality,
place the interlock connector/remote interlock back to the original position. It is
recommended to repeat this procedure each time when starting the main unit.
MANDATORY
If any of the cable connectors seems damaged or does not connect properly,
disable the device from use and contact the manufacturer for service.
C A U T IO N
Always place a dust cap to the optical fiber connector when the fiber is disconnected.
MANDATORY
If a fiber is damaged or seems twisted, contact customer service (see
rear cover page of these Instructions For Use). The fiber must be replaced
and device recalibrated by a manufacturer authorized personnel.
MANDATORY
Avoid unnecessary disconnection of the SLA optical fiber or the SLA
electrical cable from the ML6710i as it may result in changed calibration and
incorrect treatment parameters that can cause harm to the patient.
WARNING
It is recommended to use dust-free gloves when handling optical fibers. The SLA fiber
is attached to the rear panel connector as follows (presented in Figure 11):
Pay attention not to touch/contaminate the tip of the fiber
Remove the fiber connector dust cap from the rear panel by twisting it anti-clockwise
Remove the dust cap from the SLA fiber and gently push the connector inside the port in the rear panel
Twist the fiber connector nut clockwise until tightened
Disconnecting the fiber is done in a reversed order. Make sure to reattach
the dust caps to the fiber tip and the rear panel port.
Revision 9