Use Instructions
The rear panel is presented in Figure 8.
The input ports for the foot switch and interlock
connectors are located at the middle of the rear
panel. Laser emission can be turned on and
a treatment can be performed only when the
foot switch and the interlock are connected.
Fiber and SLA connector are located at the
left side. The key switch at the right side is
for switching device power on and off.
Bluetooth pairing button is located at the
top-right corner and is marked CONNECT.
It has an integrated LED which shows blue
when the device is connected to the iPad.
Do not use power inlet outside the nominal ratings specified in these
Instructions For Use and marked on the rear panel of the device.
PROHIBITED
3.4 Rear panel
The ML6710i is powered by connecting the AP0201 medical power supply to a wall socket with an AP0307
power cord. The power supply is then connected to the inlet at the right side on rear panel of the device.
The power supply automatically detects 100–240 V operating voltage and is protected against short
circuiting and overloading.
MANDATORY
Only supply voltage 100–240 VAC with frequency 50–60 Hz for
the power supply that supplies 15 VDC for the device.
Figure 8. ML6710i rear panel and connections.
Revision 9
23Product delivery package and installation